The Weiss temperature WT 64-75 test chamber of the WT series is both highly effective and economical. Its can perform cold-heat tests in extreme ranges complying with the relevant standards. An airstream conditioned to precisely the required set value is passed constantly through the test space. Components for processing the air are located in the circulating air duct installed in the rear wall of the test chamber. The symmetrical air circulation system guarantees optimum circulation. The circulating air is returned above and below the test space and flows through a heat exchanger in which it may be cooled if necessary. An electric heater warms the circulating air. An axial fan with external motor, mounted centrally in the rear test chamber wall, blows conditioned air back into the test space, providing intensive air circulation.
Applications:
The test chambers allow reproducible temperature tests to be performed in all areas of research, development, production and quality control. Regardless of the form and composition of the test specimen, efficient, even temperature control and climatisation in the test space guarantee high temperature constancy, which in turn ensures accurate test results.
Technical Specifications:
Test Space Volume - ~64ltrs
Test Space Dimensions (HxWxD mm) - 400 x 500 x 330
Temperature Range - -75 … +180°C
Interface - RS 232 C
WT 64-75 Test Chamber Specification Datasheet.pdf
The performance data refer to an ambient temperature of +25 °C